HEAT FLOW METERS

Features

  • Measurement range: 1/3 – 800 mW/m*K.
  • Automatic sample thickness measurement.
  • Open design, which allows testing whole slabs without cutting off samples (sample’s width and length are unlimited).
  • Steady State Heat flow meters with a constant temperature on both sides of the specimen: temperature fluctuation less than ±0.01 °C
  • Heat flow measured with two separate heat flow sensors.
  • Precise and fast work with excellent reproducibility.
  • Perfect for production and quality control of low thermal conductivity materials.
  • User-friendly, operation requires very little training.
  • Supplied with NIST 1450e or ERM-FC440a Certified Reference Material for the system calibration.
  • NI LabVIEW based software for measurement records, data analysis and storage, clear layout.

Custom-made solutions

Our heat flow meters can be adjusted to the customer’s individual needs or requirements. Typical custom-made solutions include:

  • Broader temperature range
  • Broader measurement range
  • Broader contact pressure range ( e.g. 0 – 20 kPa instead of standard 1 kPa)
  • Regulation of contact pressure.
  • Additional software that enables testing thermal conductivity of samples made of soft materials at various thicknesses
  • Testing smaller samples for R&D purposes, e.g. 80 x 80 mm, 250 x 250 mm.

Heat Flow Meter LM.PLUS 305

A versatile, efficient and rugged heat flow meter for industrial quality control and research & development applications.

Description

Table-mounted stand-alone unit, with 3 open sides suitable for testing larger samples without cutting. The innovative and efficient design incorporates over a decade of ongoing feedback and discussion with our industry customers and is optimized for efficient QC testing.

Uses highly reliable National Instruments electronics, software for Windows 11 is based on National Instruments LabVIEW.

The user-friendly software provides measurement records, data analysis and storage with simple and clear layout of the user interface, and simple and straightforward use.

LM.PLUS 305

  • The LM.PLUS 305 Heat Flow Meter is used for testing in accordance with standards EN 12667, ASTM C518, ISO 8301, EN 12939, DIN 52616, EN 13163, EN 13164, EN 13165.
  • In the thermal resistivity range 0.025 - 50 m 2 K/W
  • For materials including EPS, XPS, PUR, mineral wool, glass wool, wood wool, aerogels, VIP - Vacuum Insulation Panels, VIG - Vacuum Insulation Glass (VIG) and any other solid material with thermal conductivity within the specified range.
  • Calibration with supplied SRM NIST 1450e or ERM-FC440a Certified Reference Material for the system calibration.
  • For full characteristics see the technical specifications.

 

See video: LM.PLUS 305


Technical specifications:

Parameter Details
Standards EN 12667, ASTM C518, ISO 8301, EN 12939, DIN 52616, EN 13163, EN 13164, EN 13165
Type Stand alone, 3 open sides for large samples (without cutting)
Application Both QC testing and R&D, optimized for efficient QC testing
Measurement (temperature) range -5 °C to 90 °C (50 °C at default)
Temperature in the middle of the test sample 0 °C to 45 °C (85 °C on request), pre-set to 25°C at the calibration
Thermal conductivity range 0.1 – 800 mW/m*K
Thermal resistivity range 0.025 - 50 m² K/W
Materials EPS, XPS, PUR, mineral wool, glass wool, wood wool, Aerogels, VIP - Vacuum Insulation Panels, VIG - Vacuum Insulation Glass (VIG), Any material with thermal conductivity within the specified range
Measurement area (heat flux transducers area) 300 x 300 mm, customizable for R&D applications e.g. 80 x 80 mm, 250 x 250 mm
Sample dimensions (length and width) min: 300 mm, max: unlimited
Sample thickness (maximum) 100 mm
Sample thickness measurement Automatic
Sample thickness measurement precision 0.1 mm [the system permits accuracy up to 0.001 mm, but the actual measurement is limited by test sample flatness to 0.1 mm]
Measurement accuracy (thermal conductivity) ±1% to ±3% [typically better than ±1%, but may be up to ±3% in specific test specimen and conditions cases]
Repeatability 0.5%
Measurement verification NIST 1450e or ERM-FC440a Certified Reference Material
Contact load / force 2 kPa (standard), customizable in 0 - 20 kPa range
Chiller system Not Used
Cooling Air cooling
Top plate temperature Adjustable from 5°C to 50°C, pre-set at 15°C
Bottom plate temperature Adjustable from -5°C to +40°C, pre-set at 5°C
Plate thermocouples 1 to 4
Temperature Measurement resolution ±0.01 °C
Heat flux measurement 103 x 103 mm, 2 pcs. [centers of hot and cold plates]
Dimensions (L x W x H) [mm] 440 x 620 x 720
Weight [kg] Approx. 100 kg
Software features NI LabVIEW based software for measurement records, data analysis and storage, clear layout. Windows 11 OS. Computer included in the scope of delivery.
Calibration Supplied with SRM NIST 1450e or ERM-FC440a Certified Reference Material for the system calibration
Power requirements 230V / 50 Hz AC or 120 / 60 Hz AC power supply

LM.PLUS 305 with VIP + QC software

Rapid Checker for Rapid Industrial Quality Control testing of Vacuum Insulation Panels (VIPs) for large panels of unlimited width and unlimited length. QC test results: Good / Not Good.

Description

Rapid Checker developed specifically for Rapid Industrial Quality Control testing of Vacuum Insulation Panels (VIPs).

QC results: Good – measured VIP heat thermal resistivity is within acceptable limits (accepted), Not Good - measured VIP heat thermal resistivity is not within acceptable limits (rejected)

Open-frame table-mounted unit with one or more LM.PLUS 305 heat flow meters.

Uses highly reliable National Instruments electronics, software for Windows 11 is based on National Instruments LabVIEW.

The user-friendly software with simple and clear layout of the user interface, and simple and straightforward use.

LM.PLUS 305 with VIP + QC software

  • The LM.PLUS 305 VIP + QC is used for testing in accordance with standards EN 12667, ASTM C518, ISO 8301, EN 12939, DIN 52616, EN 13163, EN 13164, EN 13165.
  • In the thermal resistivity range 0.025 - 50 m2K/W.
  • Supplied with SRM NIST 1450e or ERM-FC440a Certified Reference Material for the system calibration

 

See video: LM.PLUS 305 VIP + QC

Technical specifications:

Parameter Details
Standards EN 12667, ASTM C518, ISO 8301, EN 12939, DIN 52616, EN 13163, EN 13164, EN 13165
Type open frame design with one or more LM.PLUS 305 heat flow meters [number is user specified]
Application Rapid Industrial Quality Control testing of Vacuum Insulation Panels (VIPs) for large panels; unlimited width, unlimited length. QC test results: Good / Not Good.
Measurement (temperature) range -5 °C to 90 °C (-5 °C to 50 °C at default)
Temperature in the middle of the test sample 0 °C to 45 °C (85 °C on request), pre-set to 25°C at the calibration
Thermal resistivity range 0.025 - 50 m2K/W
Materials VIP - Vacuum Insulation Panels
Measurement area (heat flux transducers area) 300 x 300 mm [for each LM.PLUS 305 VIP module]
Sample dimensions (length and width) Width: min. 300 mm, max. unlimited; Length: min. 300 mm, max. unlimited.
Sample thickness (maximum) 100 mm
Sample thickness measurement automatic
Sample thickness measurement precision 0.1 mm [the system permits accuracy up to 0.001 mm, but the actual measurement is limited by test sample flatness to 0.1 mm]
Measurement accuracy** (thermal conductivity) ±1% to ±3% [typically better than ±1%, but may be up to ±3% in specific test specimen and conditions cases]
Repeatability 0.5%
Measurement verification NIST 1450e or ERM-FC440a Certified Reference Material
Contact load / force 2 kPa (standard), customizable in 0 - 20 kPa range
Chiller system Not Used
Cooling Air cooling
Top plate temperature adjustable from 5 °C to 50 °C, pre-set at 15 °C
Bottom plate temperature adjustable from -5 °C to +40 °C, pre-set at 5 °C
Plate thermocouples 1 to 4
Temperature Measurement resolution ±0.01 °C
Heat flux measurement 103 x 103 mm, 2 pcs. [centers of hot and cold plates]
Software features NI LabVIEW based software for measurement records, data analysis and storage, clear layout. Windows 11 OS. Computer included in the scope of delivery.
Calibration Supplied with SRM NIST 1450e or ERM-FC440a Certified Reference Material for the system calibration
Power requirements 230V / 50 Hz AC or 120 / 60 Hz AC power supply

Custom-made solutions – LM.PLUS 305 VIP

Testing thermal conductivity of large samples of max. 3 m width and unlimited length.

The system can consist of one, two (as in the picture) or more measurement points. The frame size can be adjusted to individual customer needs.

The heat flow meter is open on all its sides, thus enabling measurements at the very center of a sample. This is especially suitable when testing oversized vacuum insulation panels. Such samples are more problematic to test using standard heat flow meters (whose one side is closed), because the measurement area is further from the center of the panel and more to its edge.

 


Heat Flow Meter HFM.6200

A versatile, efficient and rugged heat flow meter for industrial quality control and research & development applications that require testing large samples.

Description

Table-mounted stand-alone unit, with 3 open sides suitable for testing larger samples without cutting. The same design as with the smaller LM.PLUS 305 Heat Flow Meter, but with a larger measurement area (600 x 600 mm) for testing larger samples.

Uses highly reliable National Instruments electronics, software for Windows 11 is based on National Instruments LabVIEW.

The user-friendly software provides measurement records, data analysis and storage with simple and clear layout of the user interface, and simple and straightforward use

HFM.6200

  • The HFM.6200 Heat Flow Meter is used for testing in accordance with standards EN 12667, ASTM C518, ISO 8301, EN 12939, DIN 52616, EN 13163, EN 13164, EN 13165.
  • In the thermal resistivity range 0.025 - 50 m2K/W
  • For materials including EPS, XPS, PUR, mineral wool, glass wool, wood wool, aerogels, VIP - Vacuum Insulation Panels, VIG - Vacuum Insulation Glass (VIG) and any other solid material with thermal conductivity within the specified range.
  • Calibration with supplied SRM NIST 1450e or ERM-FC440a Certified Reference Material for the system calibration.
  • For full characteristics see the technical specifications below.

 

See video: HFM.6200

Technical specifications:

Parameter Details
Standards EN 12667, ASTM C518, ISO 8301, EN 12939, DIN 52616, EN 13163, EN 13164, EN 13165
Type stand alone, 3 open sides for large samples (without cutting)
Application both QC testing and R&D, optimized for efficient QC testing of large specimens
Measurement (temperature) range -5 °C to 90 °C (50 °C at default)
Temperature in the middle of the test sample 0 °C to 45 °C (85 °C on request), pre-set to 25°C at the calibration
Thermal conductivity range 0.1 – 800 mW/m*K
Thermal resistivity range 0.025 - 50 m2K/W
Materials EPS, XPS, PUR, mineral wool, glass wool, wood wool. Aerogels. VIP - Vacuum Insulation Panels. VIG - Vacuum Insulation Glass (VIG). Any material with thermal conductivity within the specified range.
Measurement area (heat flux transducers area) 600 x 600 mm, customizable for R&D applications e.g. 80 x 80 mm, 250 x 250 mm
Sample dimensions (length and width) min: 600 mm, max: unlimited
Sample thickness (maximum) 200 mm
Sample thickness measurement automatic
Sample thickness measurement precision 0.1 mm [the system permits accuracy up to 0.001 mm, but the actual measurement is limited by test sample flatness to 0.1 mm]
Measurement accuracy** (thermal conductivity) ±1% to ±3% [typically better than ±1%, but may be up to ±3% in specific test specimen and conditions cases]
Repeatability 0.5%
Measurement verification NIST 1450e or ERM-FC440a Certified Reference Material
Contact load / force 2 kPa (standard), customizable in 0 - 20 kPa range
Chiller system Not Used
Cooling Air cooling
Top plate temperature adjustable from 5°C to 50°C, pre-set at 15°C
Bottom plate temperature adjustable from -5°C to +40°C, pre-set at 5°C
Plate thermocouples 1 to 4
Temperature Measurement resolution ±0.01 °C
Heat flux measurement 103 x 103 mm, 2 pcs. [centers of hot and cold plates]
Dimensions (L x W x H) [mm] 850 x 1100 x 1050
Weight [kg] 270 kg
Software features NI LabVIEW based software for measurement records, data analysis and storage, clear layout. Windows 11/11 OS.
Calibration Supplied with SRM NIST 1450e or ERM-FC440a Certified Reference Material for the system calibration
Power requirements 230V / 50 Hz AC or 120 / 60 Hz AC power supply